FDA Enforcement
Class II
Terminated
AF 550 x 5ML/HR 2ML 60LO CPNB, Product code MT5060XL CPNB, Arrow AutoFuser Disposable Pain Control Pump
Recall: Z-0948-2019
·
Reported March 6, 2019
Enforcement
- Recall Number
- Z-0948-2019
- Event ID
- 81375
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Arrow International Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 6, 2019
- Initiation Date
- October 2, 2018
- Classification Date
- February 25, 2019
- Termination Date
- November 12, 2021
- Address
- 2400 Bernville Rd, Reading, PA, 19605-9607, United States
Description
AF 550 x 5ML/HR 2ML 60LO CPNB, Product code MT5060XL CPNB, Arrow AutoFuser Disposable Pain Control Pump
Reason
Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.
Code Info
74A1700874 74G1702797 74F1700146 74H1701106 74F1701964
Distribution
Nationwide distribution.
Quantity
5