FDA Enforcement Class II Terminated

Breas Ventilator Trolley with the Vivo 50/60/65 and Vivo 45LS ventilator models Article no 007384

Recall: Z-0943-2022 · Reported April 20, 2022

Enforcement

Recall Number
Z-0943-2022
Event ID
89776
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Breas Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
April 20, 2022
Initiation Date
March 21, 2022
Classification Date
April 14, 2022
Termination Date
December 13, 2024
Address
16 Esquire Rd, N/A, North Billerica, MA, 01862-2527, United States

Description

Breas Ventilator Trolley with the Vivo 50/60/65 and Vivo 45LS ventilator models Article no 007384

Reason

Ventilator mounting brackets for the Vivo 50/60/65 and Vivo 45LS ventilator models can become detached from the Breas Ventilator Trolley, this may cause the ventilator to fall from the trolley and cause injury to a person who is in the immediate proximity

Code Info

Trolleys delivered between June 11th, 2020, and December 3rd, 2021. UDI: 07321820073849

Distribution

US Nationwide

Quantity

707 units