INFINITY DUAL HEMO MCable Pod
Recall: Z-0942-2018
·
Reported March 21, 2018
Enforcement
- Recall Number
- Z-0942-2018
- Event ID
- 79209
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Draegar Medical Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 21, 2018
- Initiation Date
- October 4, 2017
- Classification Date
- March 9, 2018
- Termination Date
- June 15, 2018
- Address
- 6 Tech Dr, Andover, MA, 01810-2434, United States
Description
INFINITY DUAL HEMO MCable Pod
Reason
Dual Hemo MCable Pods with Revision Index (RI) 15 and 16, which are used with the Dr¿ger Infinity Acute Care System, may permit liquid ingress inconsistent with their IPX4 rating, which could potentially result in an incorrect measured value of invasive blood pressure.
Code Info
Have the affected revision indexes: 15 or 16 (found on the product label)
Distribution
Nationally
Quantity
2,760