FDA Enforcement Class II Ongoing

Intelerad InteleShare software, with ProViewer component

Recall: Z-0941-2026 · Reported December 24, 2025

Enforcement

Recall Number
Z-0941-2026
Event ID
97993
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
DICOM Grid, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Report Date
December 24, 2025
Initiation Date
November 18, 2025
Classification Date
December 18, 2025
Address
305 Church At North Hills St, Raleigh, NC, 27609-5177, United States

Description

Intelerad InteleShare software, with ProViewer component

Reason

Software intended to aid in diagnosing conditions, planning treatments, visualizing anatomical structures has a bug that, if all of following are met: Viewing images in InteleShare viewer; Multiplanar reconstruction applied; Manual rotation applied, could result in inaccurate length measurement tool values that could compromise diagnostic accuracy, lead to misdiagnosis or inappropriate follow-up.

Code Info

Software versions 3.24.3 - 3.25.2. Date of First Distribution: 2024-08-28. UDI-DI: B228INTELESHARE0

Distribution

US: AZ, TX, GA, NV, MO, MI, IL, OH, CT, OR, FL, NY, NM, MN, PA, AK, ND, NJ, VA, WA, NC, TN, WI, GA, AK, CA, SD, CO, MA, WV, MN, KS, MD, VT, IA, RI, AL, ID, LA, PR, NH, IN, OK, UT, NE, SC, AR, DE OUS: CA, NZ, AU, GB, BE, FR, IT, NO, SE

Quantity

924