FDA Enforcement Class II Terminated

TVC Imaging System model TVC-MC9 , for the near-infrared examination of coronary arteries in patients undergoing invasive coronary angiography The TVC Imaging System is an intravascular imaging device that simultaneously assesses vessel composition and structure using near-infrared spectroscopy (NIRS) and intravascular ultrasound (IVUS). This dual-modality instrument performs near-infrared spectroscopic analysis of the vessel to detect lipid core-containing plaques of interest (LCP) displayed in a map called a Chemogram, and simultaneously generates high resolution IVUS images that display structural details of the vessel and plaque in transverse and longitudinal views

Recall: Z-0940-2017 · Reported January 11, 2017

Enforcement

Recall Number
Z-0940-2017
Event ID
76014
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Infraredx, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 11, 2017
Initiation Date
December 15, 2016
Classification Date
January 3, 2017
Termination Date
October 9, 2018
Address
34 3rd Ave, N/A, Burlington, MA, 01803-4414, United States

Description

TVC Imaging System model TVC-MC9 , for the near-infrared examination of coronary arteries in patients undergoing invasive coronary angiography The TVC Imaging System is an intravascular imaging device that simultaneously assesses vessel composition and structure using near-infrared spectroscopy (NIRS) and intravascular ultrasound (IVUS). This dual-modality instrument performs near-infrared spectroscopic analysis of the vessel to detect lipid core-containing plaques of interest (LCP) displayed in a map called a Chemogram, and simultaneously generates high resolution IVUS images that display structural details of the vessel and plaque in transverse and longitudinal views

Reason

Near-infrared spectroscopic image (Chemogram) generated by the TVC Imaging System TVC-MC9 in some cases demonstrated an unusual appearance in that the yellow in the image (indicating the presence of lipid core plaques) was horizontally-oriented

Code Info

Serial Numbers: MC9-15013, MC9-15006, MC9-15005, MC9-15008, MC9-15012, MC9-15007 and 12718M00009 (spare laser assembly)

Distribution

Worldwide Distribution - US including MI and Internationally to South Korea

Quantity

7 units