FDA Enforcement Class I Terminated

GOPump Elastometric infusion PumpKit with/ and without accesssories all containing Dual GOPump, part number 510076: One infusion pump comprising two, 150 mL elastomeric reservoirs, each of which connects to its own catheter. Liquid flow rate is labeled at 2 mL/hr per elastomer. The Symbios Disposable Infusion Pump Kit is a disposable, self-contained infusion system utilizing an inflatable elastomeric reservoir to mechanically provide percutaneous infusion of prescribed solutions at a pre-set rate for post-operative pain management

Recall: Z-0940-2013 · Reported March 20, 2013

Enforcement

Recall Number
Z-0940-2013
Event ID
64423
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Symbios Medical Products, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 20, 2013
Initiation Date
February 25, 2013
Classification Date
March 13, 2013
Termination Date
February 13, 2014
Address
7301 Georgetown Rd., Suite 150, Indianapolis, IN, 46268, United States

Description

GOPump Elastometric infusion PumpKit with/ and without accesssories all containing Dual GOPump, part number 510076: One infusion pump comprising two, 150 mL elastomeric reservoirs, each of which connects to its own catheter. Liquid flow rate is labeled at 2 mL/hr per elastomer. The Symbios Disposable Infusion Pump Kit is a disposable, self-contained infusion system utilizing an inflatable elastomeric reservoir to mechanically provide percutaneous infusion of prescribed solutions at a pre-set rate for post-operative pain management

Reason

Received 2 complaints out of 458 distributed pumps in this lot, where flow restrictor bead became displaced, which were reported to permit fast flow of contents.

Code Info

Kit Part Number 510545-BP, lot 12-100983, 12-100984, 12-101096, 12-101125 Kit Part Number 510112-BP, lot 12-101110, 12-101103 Kit: 510401-BP lot 12-101045 Kit: 510644, lot 12-101028, 12-101097, 12-101130 GOPump Dual Infusion Pump Kit All Kits contain Pump Part Number: 510076 Lot Number: 12-100601

Distribution

Nationwide Distribution including PA, CA, OH, AL, FL, WV, TN, NJ, NC, TX, CO, NY, TX, MO, MS, SC, and GA.

Quantity

458