Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system.
Recall: Z-0939-2013
·
Reported March 20, 2013
Enforcement
- Recall Number
- Z-0939-2013
- Event ID
- 64447
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Draeger Medical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 20, 2013
- Initiation Date
- December 13, 2012
- Classification Date
- March 11, 2013
- Termination Date
- May 20, 2013
- Address
- 3135 Quarry Rd, N/A, Telford, PA, 18969-1042, United States
Description
Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system.
Reason
The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.
Code Info
Catalog Number MS20783 (Infinity MCable-Dual Hemo)
Distribution
Distributed in the states of CA, IL, and MA.
Quantity
260