Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system.

Recall: Z-0939-2013 · Reported March 20, 2013

Enforcement

Recall Number
Z-0939-2013
Event ID
64447
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Draeger Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 20, 2013
Initiation Date
December 13, 2012
Classification Date
March 11, 2013
Termination Date
May 20, 2013
Address
3135 Quarry Rd, N/A, Telford, PA, 18969-1042, United States

Description

Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system.

Reason

The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.

Code Info

Catalog Number MS20783 (Infinity MCable-Dual Hemo)

Distribution

Distributed in the states of CA, IL, and MA.

Quantity

260