FDA Enforcement
Class II
Ongoing
HCT 50mL EVA BAGS, 2 Port, Sterile, Tripple Packaged, 20 EVA BAGS/Pack, 7 Packs PER CASE. Distributed by Health Care Technology.
Recall: Z-0937-2023
·
Reported January 18, 2023
Enforcement
- Recall Number
- Z-0937-2023
- Event ID
- 91350
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Stradis Healthcare
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- January 18, 2023
- Initiation Date
- December 15, 2022
- Classification Date
- January 11, 2023
- Address
- 3600 Burwood Dr, Waukegan, IL, 60085-8399, United States
Description
HCT 50mL EVA BAGS, 2 Port, Sterile, Tripple Packaged, 20 EVA BAGS/Pack, 7 Packs PER CASE. Distributed by Health Care Technology.
Reason
Potential for the outer bag having pinhole leaks near the seal, which may result in a breach in the sterility of the kit.
Code Info
Assembly Number HCT2050, UDI-DI: M752HCT20502, Lot Number: 220130152, 221677841
Distribution
US Nationwide distribution in the states of FL, IL, IN, MA, MS, NV, OH, TN, TX, WA.
Quantity
2,779 units