FDA Enforcement
Class II
Terminated
CLARAVUE Pre-wired Electrode IP-SET N3 - PEDIATRIC-US Model: 32028776 50401-US
Recall: Z-0935-2022
·
Reported April 20, 2022
Enforcement
- Recall Number
- Z-0935-2022
- Event ID
- 89810
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Graphic Controls Acquisition Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- April 20, 2022
- Initiation Date
- March 15, 2022
- Classification Date
- April 14, 2022
- Termination Date
- July 12, 2024
- Address
- 400 Exchange St, N/A, Buffalo, NY, 14204-2064, United States
Description
CLARAVUE Pre-wired Electrode IP-SET N3 - PEDIATRIC-US Model: 32028776 50401-US
Reason
Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation.
Code Info
All lots UDI: 03700506307357
Distribution
US Nationwide distribution in the States of AR, CA, CO, FL, IN, KS, MA, MN, NY, OK, VT.
Quantity
40 units