FDA Enforcement Class II Terminated

Lumbar Extension Machine for Physical Therapy

Recall: Z-0935-2015 · Reported January 14, 2015

Enforcement

Recall Number
Z-0935-2015
Event ID
69422
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
MedX Holdings, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 14, 2015
Initiation Date
December 3, 2013
Classification Date
January 7, 2015
Termination Date
September 29, 2016
Address
839 Nw 25th Ave, N/A, Ocala, FL, 34475-5789, United States

Description

Lumbar Extension Machine for Physical Therapy

Reason

Review of DHRs revealed that HiPot, Continuity, load and friction re-testing was not completed on these devices during manufacturing.

Code Info

Serial Numbers: 01531280, 01531279, 01531278, 01531284, 01531283, 01531282, 01531281, 01531288, 01531287

Distribution

Worldwide Distribution - USA including MO, IL, ID, and MN and Internationally to Germany and South Korea.

Quantity

9