FDA Enforcement Class II Terminated

CelluTome Epidermal Harvesting System Harvester (5.0 CM) STERILE EO Rx Only Manufactured for: KCI USA, Inc. Part Number CT-H50 or CT-H25 The CelluTome Epidermal Harvesting System is intended to reproducibly cut a thin skin graft for autologous skin grafting.

Recall: Z-0933-2015 · Reported January 14, 2015

Enforcement

Recall Number
Z-0933-2015
Event ID
69961
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
KCI USA, INC.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 14, 2015
Initiation Date
November 19, 2014
Classification Date
January 7, 2015
Termination Date
September 18, 2015
Address
12930 IH 10 West, N/A, San Antonio, TX, 78249, United States

Description

CelluTome Epidermal Harvesting System Harvester (5.0 CM) STERILE EO Rx Only Manufactured for: KCI USA, Inc. Part Number CT-H50 or CT-H25 The CelluTome Epidermal Harvesting System is intended to reproducibly cut a thin skin graft for autologous skin grafting.

Reason

KCI has received reports that, in a small number of cases, the CelluTome Harvester (a component of the CelluTome Epidermal Harvesting System), blades have the potential to drift during shipping and could prematurely protrude into openings of harvester top plate.

Code Info

CT-H25 lot numbers: 3249980000, 3249990000, 3250010000, 2537872, 2556866; CT-H50 lot numbers: 3249660000, 3249840000, 3250050000, 2498886, 2510874, 2537879, 2545871

Distribution

Worldwide Distribution - USA (Nationwide) and Internationally to Australia, Austria, Canada, France, Germany, Ireland, Italy, Isreal, Netherlands, South Africa, Spain, Sweden, Switerland, Turkey, UAE, and United Kingdom.

Quantity

1,139/5 Packs