FDA Enforcement
Class II
Terminated
Proteus 235, Pencil Beam Scanning version PTS-6.4.11.X The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck.
Recall: Z-0931-2018
·
Reported March 14, 2018
Enforcement
- Recall Number
- Z-0931-2018
- Event ID
- 79228
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Ion Beam Applications S.A.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 14, 2018
- Initiation Date
- October 19, 2017
- Classification Date
- March 8, 2018
- Termination Date
- April 2, 2019
- Address
- Chemin du Cyclotron, 3, N/A, Louvain La Neuve, N/A, N/A, Belgium
Description
Proteus 235, Pencil Beam Scanning version PTS-6.4.11.X The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck.
Reason
Range for the same energy could be different depending if the layer is the first layer of the irradiation plan or not.
Code Info
PAT.003 (KR), PAT.006 (US)
Distribution
US Distribution to the state of :FL
Quantity
3