FDA Enforcement Class II Ongoing

Atellica CH Fructosamine (Fruc), Siemens Material Number 11097637

Recall: Z-0930-2022 · Reported April 20, 2022

Enforcement

Recall Number
Z-0930-2022
Event ID
89740
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Siemens Healthcare Diagnostics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 20, 2022
Initiation Date
February 9, 2022
Classification Date
April 14, 2022
Address
511 Benedict Ave, N/A, Tarrytown, NY, 10591-5005, United States

Description

Atellica CH Fructosamine (Fruc), Siemens Material Number 11097637

Reason

Siemens Healthcare Diagnostics Inc. has confirmed the potential for reagent carryover impacting quality control (QC), patient samples, and calibrator results.

Code Info

UDI: 00630414595580; Lot Numbers: 211359 211451

Distribution

Global Distribution. US Nationwide.

Quantity

199