FDA Enforcement Class II Terminated

Proteus 235, graphite block 8

Recall: Z-0930-2018 · Reported March 14, 2018

Enforcement

Recall Number
Z-0930-2018
Event ID
79227
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ion Beam Applications S.A.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 14, 2018
Initiation Date
October 17, 2017
Classification Date
March 8, 2018
Termination Date
March 29, 2019
Address
Chemin du Cyclotron, 3, N/A, Louvain La Neuve, N/A, N/A, Belgium

Description

Proteus 235, graphite block 8

Reason

IBA identified incorrect screw holes depth in graphite block 8 with respect to the degrader

Code Info

PAT.115, SAT.116

Distribution

WA

Quantity

3