FDA Enforcement Class II Terminated

Dexcom Receiver, REF: MT22719, (Black, Pin, Blue), Rx Only, containing software version SW10617 and revision 4.0.1.048 - Product Usage: A non-adjunctive invasive glucose sensor is intended to determine glucose levels, and the direction and rate of change of glucose levels in people with diabetes. The device is indicated to replace information obtained from standard blood glucose monitoring devices to make diabetes-related treatment decisions. The device also provides historical glucose information, facilitating long-term therapy adjustments.

Recall: Z-0929-2020 · Reported February 12, 2020

Enforcement

Recall Number
Z-0929-2020
Event ID
84290
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Dexcom Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 12, 2020
Initiation Date
July 12, 2019
Classification Date
January 31, 2020
Termination Date
December 15, 2022
Address
6340 Sequence Dr, N/A, San Diego, CA, 92121-4356, United States

Description

Dexcom Receiver, REF: MT22719, (Black, Pin, Blue), Rx Only, containing software version SW10617 and revision 4.0.1.048 - Product Usage: A non-adjunctive invasive glucose sensor is intended to determine glucose levels, and the direction and rate of change of glucose levels in people with diabetes. The device is indicated to replace information obtained from standard blood glucose monitoring devices to make diabetes-related treatment decisions. The device also provides historical glucose information, facilitating long-term therapy adjustments.

Reason

It has been reported that use of the mobile receiver with software version SW10617 rev 4.0.1.048 have reported: 1. the receiver becoming stuck on initialization screen when powering on. This will cause patients not to be able to receive glucose values or alerts; 2. Reinitialization of the receiver without user interaction. If this occurs, the receiver will either reboot and operate normally or require the user to press the Select button on the receiver in order to resume normal operation. In this situation, the user is provided an audio and vibratory alert every 5 minutes. Until the select button is pressed, patients will not receive glucose values or alerts.

Code Info

All serial numbers of the G5 Mobile Receivers with software version SW10617 and software revision 4.0.1.048 Receiver Part Number/Reciver Kit Number/GUDID registered GTINS/ Receiver Kit Type: Dexcom G5 Mobile Receiver, MT22179 (BLK) Rev 013 and Rev 014 Receiver (mg/dL), Black STK-GF-001/ 10386270000238 - Starter Kit STK-RF-001/20386270000235 - Retail Starter Kit STR-GF-001/30386270000232 - Replacement Receiver Kit STK-GF-IUO/NA - Investigational Use STK-MC-001/00386270000606 - Medicare Starter Kit Dexcom G5 Mobile Receiver, MT22719 (PNK) Rev 012 and 013, Receiver (mg/dL), Pink STK-GF-PNK/ 30386270000263 - Starter Kit STK-RF-PNK/10386270000269 - Retail Starter Kit STR-GF-PNK/20386270000266 - Replacement Receiver Kit Dexcom G5 Mobile Receiver, MT22719 (BLU) Rev 012 and 013, Receiver (mg/dL), Blue STK-GF-BLU/ 30386270000270 - Starter Kit STK-RF-BLU/10386270000276 - Retail Starter Kit STR-GF-BLU/20386270000273 - Replacement Receiver Kit

Distribution

US Nationwide distribution.

Quantity

112,415 receivers