FDA Enforcement Class II Terminated

DIASTAT anti-Nuclear antibody (ANA) / DIASTAT ANA ELISA. Catalog Number: FANA200. Product Usage: The DIASTAT¿ anti-nuclear antibody (ANA) test is a qualitative enzyme-linked immunosorbent assay (ELISA) for the detection of ANAs in human serum or EDTA plasma. It detects ANAs against Sm, Sm/RNP, Ro (SS-A), La (SS-B), Scl- 70, Jo-1, dsDNA, histone and centromere antigens.The test may be used to screen out samples which are negative for all ANAs. Samples that give a positive test result should be further tested to identify the antigen-specific antibody or antibodies present. Quantitative/qualitative ELISA kits are available for the individual detection of Sm, Sm/RNP, Ro (SS-A), La (SS-B), dsDNA and centromere ANAs. Qualitative kits are available for the individual detection of Scl-70 and Jo-1. ANA detection represents one parameter in a multicriterion diagnostic process.

Recall: Z-0929-2018 · Reported March 14, 2018

Enforcement

Recall Number
Z-0929-2018
Event ID
79183
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Euro Diagnostica AB
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
March 14, 2018
Initiation Date
October 15, 2015
Classification Date
March 8, 2018
Termination Date
March 29, 2019
Address
Box 50117, Lundavagen 151, Malmo, N/A, N/A, Sweden

Description

DIASTAT anti-Nuclear antibody (ANA) / DIASTAT ANA ELISA. Catalog Number: FANA200. Product Usage: The DIASTAT¿ anti-nuclear antibody (ANA) test is a qualitative enzyme-linked immunosorbent assay (ELISA) for the detection of ANAs in human serum or EDTA plasma. It detects ANAs against Sm, Sm/RNP, Ro (SS-A), La (SS-B), Scl- 70, Jo-1, dsDNA, histone and centromere antigens.The test may be used to screen out samples which are negative for all ANAs. Samples that give a positive test result should be further tested to identify the antigen-specific antibody or antibodies present. Quantitative/qualitative ELISA kits are available for the individual detection of Sm, Sm/RNP, Ro (SS-A), La (SS-B), dsDNA and centromere ANAs. Qualitative kits are available for the individual detection of Scl-70 and Jo-1. ANA detection represents one parameter in a multicriterion diagnostic process.

Reason

Complaint investigation concluded that although product quality requirements stated in the instructions for use were met at lot release, the high background in the ELISA plate, the strong IgM conjugate, and the low reference control led to an increased risk of equivocal and/or false positive sample test results for the two FANA200 kit lots SS 1500 and SS 2009.

Code Info

UDI: 07340058410486. Lot numbers: SS 1500 and SS 2009.

Distribution

Worldwide Distribution and US. Nationwide

Quantity

1090