EndoVive 3s Low Profile Balloon Kits Part Number: M00548680 (XMD P/N 70-0050-318) The EndoVive 3s Low Profile Balloon is intended to provide gastric access through a gastrointestinal stoma tract for enteral feeding, medication administration and gastric decompression. The EndoVive 3s Low Profile Balloon is indicated for use in adult and pediatric patients who require long-term feeding, are unable to tolerate oral feeding, are at low risk for aspiration or require gastric decompression and/or medication delivery directly into the stomach.
Enforcement
- Recall Number
- Z-0929-2016
- Event ID
- 73024
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Xeridiem Mediem Medical Devices Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 2, 2016
- Initiation Date
- December 23, 2015
- Classification Date
- February 25, 2016
- Termination Date
- October 25, 2016
- Address
- 4700 S Overland Dr, N/A, Tucson, AZ, 85714-3430, United States
Description
EndoVive 3s Low Profile Balloon Kits Part Number: M00548680 (XMD P/N 70-0050-318) The EndoVive 3s Low Profile Balloon is intended to provide gastric access through a gastrointestinal stoma tract for enteral feeding, medication administration and gastric decompression. The EndoVive 3s Low Profile Balloon is indicated for use in adult and pediatric patients who require long-term feeding, are unable to tolerate oral feeding, are at low risk for aspiration or require gastric decompression and/or medication delivery directly into the stomach.
A feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric contents to leak out of the device in some instances.
Part Number: M00548680 (XMD P/N 70-0050-318) Lot Number: 928919 929709
Nationwide distribution to MA only.
78