FDA Enforcement
Class II
Terminated
BIOMERIEUX VIDAS 3, Compact immunoanalyzer with full traceability and automation
Recall: Z-0928-2020
·
Reported February 12, 2020
Enforcement
- Recall Number
- Z-0928-2020
- Event ID
- 84464
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Biomerieux Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- February 12, 2020
- Initiation Date
- October 22, 2019
- Classification Date
- January 31, 2020
- Termination Date
- August 6, 2024
- Address
- 595 Anglum Rd, N/A, Hazelwood, MO, 63042-2320, United States
Description
BIOMERIEUX VIDAS 3, Compact immunoanalyzer with full traceability and automation
Reason
Following Customers complaints, investigations have been initiated on potentially false results obtained on VIDAS 3 with an expired calibration. Indeed, the calibrations of assays were valid in the calibration menu whereas in fact the calibrations were expired and no alarm displayed to warn the users about the expiration date of the calibrations. Software computes results of analysis for assays with expired calibrations. The anomaly is due to an incorrect update of the calibration status by the software.
Code Info
software versions 1.2 and higher
Distribution
US
Quantity
738 units