Cable remote control JUC, BUTTON, Material no.1239823, Serial number: to (incl.) 100143048. Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.
Enforcement
- Recall Number
- Z-0928-2015
- Event ID
- 69927
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Trumpf Medical Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 14, 2015
- Initiation Date
- December 3, 2014
- Classification Date
- January 6, 2015
- Termination Date
- August 23, 2016
- Address
- 1046 Legrand Blvd, N/A, Charleston, SC, 29492-7672, United States
Description
Cable remote control JUC, BUTTON, Material no.1239823, Serial number: to (incl.) 100143048. Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes
The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188.
Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.
3303 total all devices