FDA Enforcement Class II Terminated

i-STAT DE handheld data processing module for clinical use, Software Version 2.8, List Number:08K46-01115200 (UDI: 00054749001255)

Recall: Z-0926-2018 · Reported March 14, 2018

Enforcement

Recall Number
Z-0926-2018
Event ID
79150
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Abbott Point Of Care Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
March 14, 2018
Initiation Date
September 12, 2017
Classification Date
March 8, 2018
Termination Date
August 23, 2018
Address
400 College Rd E, N/A, Princeton, NJ, 08540-6607, United States

Description

i-STAT DE handheld data processing module for clinical use, Software Version 2.8, List Number:08K46-01115200 (UDI: 00054749001255)

Reason

Issues resulting from upgrade to software version 2.8: (1) Location, operator, stored patient lists will not update, and (2) Customized Reference Ranges, Action Ranges, and Custom Reportable Ranges are reset to factory default values. No erroneous results are generated as a result of this issue.

Code Info

All units that have been upgraded to software version 2.8

Distribution

Distributed in 48 States: AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, the District of Columbia, Australia, Italy, Sweden, and Canada.

Quantity

894