i-STAT DE handheld data processing module for clinical use, Software Version 2.8, List Number:08K46-01115200 (UDI: 00054749001255)
Enforcement
- Recall Number
- Z-0926-2018
- Event ID
- 79150
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Abbott Point Of Care Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- March 14, 2018
- Initiation Date
- September 12, 2017
- Classification Date
- March 8, 2018
- Termination Date
- August 23, 2018
- Address
- 400 College Rd E, N/A, Princeton, NJ, 08540-6607, United States
Description
i-STAT DE handheld data processing module for clinical use, Software Version 2.8, List Number:08K46-01115200 (UDI: 00054749001255)
Issues resulting from upgrade to software version 2.8: (1) Location, operator, stored patient lists will not update, and (2) Customized Reference Ranges, Action Ranges, and Custom Reportable Ranges are reset to factory default values. No erroneous results are generated as a result of this issue.
All units that have been upgraded to software version 2.8