FDA Enforcement Class II Ongoing

Pediatric Craniotome Ref:CRANI-P-G1

Recall: Z-0924-2023 · Reported January 18, 2023

Enforcement

Recall Number
Z-0924-2023
Event ID
91231
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
The Anspach Effort, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 18, 2023
Initiation Date
December 7, 2022
Classification Date
January 6, 2023
Address
4500 Riverside Dr, Palm Beach Gardens, FL, 33410-4235, United States

Description

Pediatric Craniotome Ref:CRANI-P-G1

Reason

Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury

Code Info

GTIN: 00845384016403 All lots

Distribution

US Nationwide and OUS Foreign: Impacted Foreign Countries Australia Bangladesh China Hongkong India Indonesia Japan Korea (Republic of) Malaysia New Zealand Philippines Singapore Taiwan, Province of China Thailand Angola Austria Belgium Croatia Cyprus Czech Republic Denmark Equatorial Guinea Finland France & French Guiana Georgia Germany Greece Hungary Israel Italy Kenya Kuwait Lebanon Lithuania Macedonia Mauritius Morocco Netherlands Poland Portugal Saudi Arabia Slovakia South Africa Spain Sweden Switzerland Turkey United Arab Emirates United Kingdom of Great Britain and Northern Ireland Argentina Bolivia Brazil Chile Colombia Costa Rica Jamaica Mexico Panama Peru Canada

Quantity

6059 units in total