FDA Enforcement Class II Terminated

ABS800 Chemistry Analyzer Part Number BA81F-PA00001, Containing Gas spring YQ-8/18-90-272-180N (Material Code M6T-010001--- (front spring)) and Gas spring YQ6/15-84-235(B-B)-80N (033-000090-00 (rear spring))

Recall: Z-0921-2020 · Reported February 5, 2020

Enforcement

Recall Number
Z-0921-2020
Event ID
84497
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Mindray DS USA, Inc. dba Mindray North America
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 5, 2020
Initiation Date
December 9, 2019
Classification Date
January 30, 2020
Termination Date
September 17, 2020
Address
800 Macarthur Blvd, N/A, Mahwah, NJ, 07430-2001, United States

Description

ABS800 Chemistry Analyzer Part Number BA81F-PA00001, Containing Gas spring YQ-8/18-90-272-180N (Material Code M6T-010001--- (front spring)) and Gas spring YQ6/15-84-235(B-B)-80N (033-000090-00 (rear spring))

Reason

The gas spring on the front and rear protective cover may fail. There is a low risk of injury to the user resulting from failure of the gas spring. Injury may include impact to the arms, hands or fingers.

Code Info

Serial Numbers W2-6C000001, W2-6C000002, W2-6C000003, W2-6C000004, W2-6C000005, XP-5B000611T

Distribution

The products were distributed to the following US states: MI and TX.

Quantity

6