FDA Enforcement Class II Ongoing

Implant Epikut S Plus CM 16-Degrees 4,5 X 18,0 mm, REF: ILM 4518N

Recall: Z-0920-2025 · Reported January 22, 2025

Enforcement

Recall Number
Z-0920-2025
Event ID
95983
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
SIN SISTEMA DE IMPLANTE NACIONAL S.A
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 22, 2025
Initiation Date
September 1, 2024
Classification Date
January 15, 2025
Address
Jardim Analia Franco, Rua Soldador Ocimar Guimaraes da Silva 421, Sao Paulo, N/A, Brazil

Description

Implant Epikut S Plus CM 16-Degrees 4,5 X 18,0 mm, REF: ILM 4518N

Reason

Dental implant box incorrectly labels the diameter as 4,5 X 18,0 mm, but the internal label correctly labels the diameter as 4,0 X 18,0 mm; if the dentist drills the bone with the larger diameter stated on the box, the smaller diameter implant, will become loose and will not fit, which may necessitate an additional surgery.

Code Info

UDI-DI: 07899995273437, Lot: X010469757, Expiration: 2028/01

Distribution

US distribution to states of: KY, CO, CA, NY, ID, NJ, FL, TX, KY, IL, OH; AND OUS (foreign) country of Brazil.

Quantity

123 (US 98, OUS 25)