FDA Enforcement
Class II
Ongoing
OsteoCove Putty, 1.25cc, REF 56920013 and Cove Putty, 1.25cc REF 02-9200-01.
Recall: Z-0919-2025
·
Reported January 22, 2025
Enforcement
- Recall Number
- Z-0919-2025
- Event ID
- 95978
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- IsoTis OrthoBiologics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- January 22, 2025
- Initiation Date
- November 27, 2024
- Classification Date
- January 15, 2025
- Address
- 2 Goodyear Ste A, Irvine, CA, 92618-2052, United States
Description
OsteoCove Putty, 1.25cc, REF 56920013 and Cove Putty, 1.25cc REF 02-9200-01.
Reason
Hydration issue resulting in the product being less cohesive and moldable than intended due to an issue with the polymer.
Code Info
OsteoCove Putty, REF 56920013, UDI-DI: 10889981309022, Lot Numbers: 180054, Exp. 2025-08-28; 180054R18, Exp. 2025-08-28; 179604R18, Exp. 2024-12-28; 180484, Exp. 2026-02-28; 180370, Exp. 2026-01-28; 180299, Exp. 2025-11-28; 180075, Exp. 2025-08-28; and Cove Putty, REF 02-9200-013, UDI-DI: 10889981308933, Lot Number: 180075R18, Exp. 2025-08-28.
Distribution
US Nationwide distribution in the states of AL, CA, MD. MN, MS, NC, NV, OH, OK, TX.
Quantity
145