Lotus TAVR 23mm, Transcatheter Aortic Valve Prosthesis Premounted on Delivery System; Sterile Material number H749LTV230, Catalog Number LTV23; Product Usage: The Lotus Valve System is intended to improve aortic valve function for symptomatic subjects with severe calcific aortic stenosis (aortic valve area [AVA] of <1.0 cm2 or AVA index of <0.6 em /m2) who are at extreme or high risk for standard surgical valve replacement.
Enforcement
- Recall Number
- Z-0915-2015
- Event ID
- 69931
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Boston Scientific Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 21, 2015
- Initiation Date
- November 19, 2014
- Classification Date
- January 13, 2015
- Termination Date
- March 26, 2015
- Address
- 160 Knowles Dr, Los Gatos, CA, 95032-1828, United States
Description
Lotus TAVR 23mm, Transcatheter Aortic Valve Prosthesis Premounted on Delivery System; Sterile Material number H749LTV230, Catalog Number LTV23; Product Usage: The Lotus Valve System is intended to improve aortic valve function for symptomatic subjects with severe calcific aortic stenosis (aortic valve area [AVA] of <1.0 cm2 or AVA index of <0.6 em /m2) who are at extreme or high risk for standard surgical valve replacement.
Lotus valve became unlocked during release from the delivery system. This may lead to percutaneous insertion of another valve, resulting in a Valve in Valve (ViV) implantation. If this occurs, the first valve could embolize and it may be necessary to convert the patient to surgery.
Serial numbers: 14113043, 14118011, 14118012, 14119047, 14119048, 14119049, 14119050, 14119051, 14120027, 14120028, 14120029, 14120030, 14120031, 14121035, 14121036, 14121037, 14121038, 14121039, 14121040, 14121041, 14125040, 14125041, 14125042, 14125043, 14125044, 14125045, 14125046, 14125047, 14125048, 14125049, 14127073, 14127077, 14127078, 14127079, 14127080, 14127081, 14127082, 14127083, 14127084, 14127085, 14127086, 14141044, 14141045, 14141046, 14141047, 14141048, 14141049, 14141050, 14141051, 14141052, 14141053, 14148034, 14148035, 14148036, 14148037, 14148039, 14153054, 14153055, 14153057, 14153058, 14153059, 14153075, 14153076, 14153077, 14153078, 14153079, 14157003, 14157015, 14157016, 14157017, 14157019, 14157020, 14157021, 14157022, 14157023, 14178024, 14178025, 14178026, 14178027, 14178028. Nov. 15, 2014 to Jan. 24, 2015.
Distributed only in the countries of Finland, France, Germany, Great Britain, Italy, Norway, Spain, Sweden, and Switzerland.
278 units, all sizes