FDA Enforcement Class II Terminated

VITROS Chemistry Products DAT Performance Verifier IV in vitro diagnostic

Recall: Z-0914-2018 · Reported March 14, 2018

Enforcement

Recall Number
Z-0914-2018
Event ID
79155
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ortho-Clinical Diagnostics
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 14, 2018
Initiation Date
September 22, 2017
Classification Date
March 6, 2018
Termination Date
August 24, 2020
Address
1000 Lee Road, N/A, Rochester, NY, 14606, United States

Description

VITROS Chemistry Products DAT Performance Verifier IV in vitro diagnostic

Reason

This product is not meeting the current storage labeling for unopened VITROS DAT Performance Verifiers causing customers to obtain lower than expected VITROS BENZ results when using this control.

Code Info

Catalog# 6802330 Lot# L5619

Distribution

Worldwide distribution: US Nationwide (including Puerto Rico), Bermuda, Australia, Brazil, Canada, Chile, Colombia, India, Japan, Mexico, Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain, Poland, and Russia.

Quantity

196