Nextra Hammertoe Correction System 3.5 Middle 3.2 Proximal Kit SKU: NX-3532K Orthopedic surgical instrument
Recall: Z-0912-2022
·
Reported April 20, 2022
Enforcement
- Recall Number
- Z-0912-2022
- Event ID
- 89764
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Nextremity Solutions
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- April 20, 2022
- Initiation Date
- February 10, 2022
- Classification Date
- April 12, 2022
- Termination Date
- October 10, 2024
- Address
- 1195 Polk Dr, N/A, Warsaw, IN, 46582-8602, United States
Description
Nextra Hammertoe Correction System 3.5 Middle 3.2 Proximal Kit SKU: NX-3532K Orthopedic surgical instrument
Reason
Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery
Code Info
Lot Number: 168125017A UDI: 00817701020042
Distribution
Distribution US Nationwide, Japan, and the Netherlands
Quantity
285 units