FDA Enforcement Class II Terminated

Olympus, Model No. XCF-Q160ALE, flexible scope, large diameter, video colonoscope

Recall: Z-0910-2016 · Reported March 2, 2016

Enforcement

Recall Number
Z-0910-2016
Event ID
73088
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Integrated Medical Systems Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 2, 2016
Initiation Date
January 20, 2016
Classification Date
February 25, 2016
Termination Date
October 13, 2016
Address
3200 2nd Ave N, N/A, Birmingham, AL, 35222-1212, United States

Description

Olympus, Model No. XCF-Q160ALE, flexible scope, large diameter, video colonoscope

Reason

If excessive pressure is applied to the distal end of the endoscope, the internal coil pipe assembly at the distal end of the endoscope can become detached and angulation may not operate as intended.

Code Info

Serial Number 240T024

Distribution

Distributed US (nationwide) and in Australia.

Quantity

1 instrument