FDA Enforcement
Class II
Terminated
SIEMENS brand ARTISTE syngo RT Therapist (RTT) version 4.3, ARTISTE Model, Part No. 08162815, Digital linear accelerators, MOSAIQ OIS system, Distributed by SIEMENS Healthcare IM CR RO, CR Radiation Oncology, Manufactured by SIEMENS AG, MEDICAL SOLUTIONS, GERMANY.
Recall: Z-0907-2013
·
Reported March 13, 2013
Enforcement
- Recall Number
- Z-0907-2013
- Event ID
- 64214
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Medical Solutions USA, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 13, 2013
- Initiation Date
- January 17, 2013
- Classification Date
- March 5, 2013
- Termination Date
- March 5, 2013
- Address
- 4040 Nelson Ave, N/A, Concord, CA, 94520-1200, United States
Description
SIEMENS brand ARTISTE syngo RT Therapist (RTT) version 4.3, ARTISTE Model, Part No. 08162815, Digital linear accelerators, MOSAIQ OIS system, Distributed by SIEMENS Healthcare IM CR RO, CR Radiation Oncology, Manufactured by SIEMENS AG, MEDICAL SOLUTIONS, GERMANY.
Reason
Siemens became aware of a potential safety issue when using the "In-session resumption" feature of the syngo RT Therapist (RTT) 4.3 software. The RTT will send again the previously completed beam or segment for delivery to the control console, thus repeating the already delivered Monitor Units of the same beam or segment.
Code Info
All codes/units of ARTISTE model.
Distribution
5 devices, all located outside of the United States in Germany.
Quantity
5 units