FDA Enforcement Class II Terminated

Separation Technology Creamatocrit Plus, Catalog Number: 100-146

Recall: Z-0905-2022 · Reported April 20, 2022

Enforcement

Recall Number
Z-0905-2022
Event ID
89829
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stanbio Laboratory, LP
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 20, 2022
Initiation Date
March 16, 2022
Classification Date
April 11, 2022
Termination Date
January 2, 2025
Address
1261 N Main St, N/A, Boerne, TX, 78006-3014, United States

Description

Separation Technology Creamatocrit Plus, Catalog Number: 100-146

Reason

The device was being marketed as an IVD without a 510(k). Therefore, performance characteristics of these breast milk analyzers have not been adequately validated. An erroneous result could lead to sub-optimal nutrition supplementation for patients (infants)

Code Info

Sold between 2019 and 2021

Distribution

US Nationwide Distribution CA, TX, OK, SD, IL, NV, MO, FL, WA, KY, WI, PA, NC, GA, MD, MN

Quantity

50