FDA Enforcement Class II Ongoing

Liquid Assayed Specific Protein- In vitro diagnostic use, in the quality control of serum on clinical chemistry and immunoassay systems Catalog Number: PS2682 Control Level 1

Recall: Z-0903-2023 · Reported January 11, 2023

Enforcement

Recall Number
Z-0903-2023
Event ID
91225
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Randox Laboratories Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
January 11, 2023
Initiation Date
November 22, 2022
Classification Date
January 5, 2023
Address
55 Diamond Road, N/A, Crumlin (North), N/A, N/A, Ireland

Description

Liquid Assayed Specific Protein- In vitro diagnostic use, in the quality control of serum on clinical chemistry and immunoassay systems Catalog Number: PS2682 Control Level 1

Reason

Concentration of Rheumatoid Factor has decreased in Specific Protein Control Level 1-3 causing delay in reporting results due to Quality Controls running low outside range

Code Info

GTIN: 05055273204896 Lot Number/Exp. Date: 584LPC 28th Jun 23; 600LPC 28th Nov 23; 611LPC 28th Sep 24;

Distribution

US Nationwide distribution in the states of CA, FL, OH,ME, MO, NC, NJ, PA, TN.

Quantity

146 units US