FDA Enforcement Class II Ongoing

Cardiohelp-i Transport Guard with Transport Guard (701072525): is a blood oxygenation and carbon dioxide removal system used to pump blood through the extracorporeal bypass circuit for circulatory and/or pulmonary support for periods appropriate to cardiopulmonary bypass (up to six hours). Product Code/REF Number: 701072780 (Cardiohelp-i)

Recall: Z-0897-2022 · Reported April 13, 2022

Enforcement

Recall Number
Z-0897-2022
Event ID
89745
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Maquet Medical Systems USA
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 13, 2022
Initiation Date
February 18, 2022
Classification Date
April 8, 2022
Address
45 Barbour Pond Dr, Wayne, NJ, 07470-2094, United States

Description

Cardiohelp-i Transport Guard with Transport Guard (701072525): is a blood oxygenation and carbon dioxide removal system used to pump blood through the extracorporeal bypass circuit for circulatory and/or pulmonary support for periods appropriate to cardiopulmonary bypass (up to six hours). Product Code/REF Number: 701072780 (Cardiohelp-i)

Reason

Potential for a loose cylindrical pin on the Transport Guard that may result in the jamming of the locking mechanism causing the Transport Guard to not open or, if the locking pin is fully disengaged from the assembly, a full disassembly of the Transport Guard may occur, resulting in a delay of support if a product exchange is necessary.

Code Info

Serial Numbers for the Cardiohelp-i (Part Number 701072780)as the Transport Guard (701072525) is packaged with the Cardiohelp-i and is not labelled individually Serial Numbers: 90413945, 90414225, 90414226, 90414227, 90414228, 90414229, 90414232, 90414233, 90414234, 90414235, 90414236, 90414237, 90414283, 90414284, 90414285, 90414286, 90414287, 90414288, 90414289, 90414292, 90414293, 90414294, 90414295, 90414296, 90414297, 90414309, 90414310, 90414311, 90414312, 90414317, 90414318, 90414319. UDI: 04058863074863 (Cardiohelp-i)

Distribution

US Nationwide Distribution: FL, KY, NH, NJ, NY, PA, TX, VA Foreign: Belgium, Canada, Germany, Hungary, Luxemburg, Malaysia, Mexico, Poland, Slovakia, South Africa, South Korea, and Taiwan.

Quantity

US: 32 units; OUS 30 units