FDA Enforcement Class II Terminated

Sensei X Robotic Catheter System, model numbers 02057 and 08257. - Set up joint (SUJ)mechanical clamp that secures the SUJ to the catheter laboratory table. Product Usage: The Sensei X Robotic Catheter System is an electromechanical software controlled system designed to facilitate navigation and control of Hansen Medicals robotically steerable catheters for collecting electrophysiological data within the heart atria with electro-anatomic mapping and recording systems.

Recall: Z-0896-2015 · Reported January 7, 2015

Enforcement

Recall Number
Z-0896-2015
Event ID
69855
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Hansen Medical Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 7, 2015
Initiation Date
December 4, 2014
Classification Date
December 30, 2014
Termination Date
September 3, 2015
Address
800 E Middlefield Rd, N/A, Mountain View, CA, 94043-4030, United States

Description

Sensei X Robotic Catheter System, model numbers 02057 and 08257. - Set up joint (SUJ)mechanical clamp that secures the SUJ to the catheter laboratory table. Product Usage: The Sensei X Robotic Catheter System is an electromechanical software controlled system designed to facilitate navigation and control of Hansen Medicals robotically steerable catheters for collecting electrophysiological data within the heart atria with electro-anatomic mapping and recording systems.

Reason

The Set Up Joint subassembly of the Sensei X Robotic Catheter System may become loose relative to the operating table with the potential for the device to fall causing, uncontrolled catheter movement. cardiac tampenade, delayed procedure, or crush injury to the patient and/or user.

Code Info

Model numbers 02057 and 08257, all installed Sensei X Robotic systems are subject to this action.

Distribution

Worldwide Distribution - US Nationwide and the countries of Italy, Czech Republic, Russia, Spain, Singapore, France, Germany, Denmark, Netherlands and Australia.

Quantity

59 USA, 40 OUS