FDA Enforcement Class II Terminated

Firebird Spinal Fixation System, Parallel Rod Connector (52-6800) and Axial Rod Connector (52-6700). Product Usage: The connector assemblies allow for rod components to be connected side-to-side, or end-to-end and eliminate the need of rebuilding the entire system with a longer rod in order to include the new vertebral levels.

Recall: Z-0896-2013 · Reported March 6, 2013

Enforcement

Recall Number
Z-0896-2013
Event ID
64265
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Orthofix, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 6, 2013
Initiation Date
February 1, 2013
Classification Date
February 28, 2013
Termination Date
July 1, 2013
Address
3451 Plano Pkwy, N/A, Lewisville, TX, 75056-9453, United States

Description

Firebird Spinal Fixation System, Parallel Rod Connector (52-6800) and Axial Rod Connector (52-6700). Product Usage: The connector assemblies allow for rod components to be connected side-to-side, or end-to-end and eliminate the need of rebuilding the entire system with a longer rod in order to include the new vertebral levels.

Reason

There is a possibility that the Set Screw (p/n 90-2923) sub-component of the Parallel Rod Connector Assembly (p/n 52-6800) & Axial Rod Connector (p/n 52-6700) may crack during final tightening. Orthofix Inc received six (6) customer complaints related to set screw cracking between November 15, 2010 and December 2012.

Code Info

Catalog Numbers: 52-6800 & 52-6700; All Lot Numbers

Distribution

Worldwidewide Distribution - USA Nationwide including Puerto Rico and the countries of: Australia, Lebanon, Germany, Spain, Sweden and Israel.

Quantity

565 total (287 PN 52-6800 and 278 PN 52- 6700)