FDA Enforcement Class II Terminated

CS Elite Fastpack, 225 mL, 20¿ Res Autotransfusion Device - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for gravity reinfusion into the patient or to the arterial line of an extracorporeal circuit for reinfusion into the patient.

Recall: Z-0894-2020 · Reported February 5, 2020

Enforcement

Recall Number
Z-0894-2020
Event ID
84183
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Haemonetics Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 5, 2020
Initiation Date
October 21, 2019
Classification Date
January 29, 2020
Termination Date
November 6, 2020
Address
400 Wood Rd, N/A, Braintree, MA, 02184-2412, United States

Description

CS Elite Fastpack, 225 mL, 20¿ Res Autotransfusion Device - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for gravity reinfusion into the patient or to the arterial line of an extracorporeal circuit for reinfusion into the patient.

Reason

Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.

Code Info

Catalog number/FG Item Number: CSE-FP-225F 1. Specific Lots with Bowl Cracks (452 Distributed) Lot numbers: 1901003HTT 1810008HTT 2. Missing instructions to check for cracks in IFU All products/lost manufactured between October 2014 and October 2019.

Distribution

Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.

Quantity

1092