FDA Enforcement Class II Ongoing

Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,

Recall: Z-0887-2026 · Reported December 17, 2025

Enforcement

Recall Number
Z-0887-2026
Event ID
97864
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Bard Peripheral Vascular Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
N/A
Report Date
December 17, 2025
Initiation Date
November 6, 2025
Classification Date
December 5, 2025
Address
1625 W 3rd St Bldg 1, 850 W. Rio Salado Prkwy., Tempe, AZ, 85281-2438, United States

Description

Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,

Reason

Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels.

Code Info

UDI/Lot/Expiration: (01)10885403108365(17)271130(10)0001623320/0001623320/20271130; (01)10885403108365(17)271130 (10)0001623069/0001623069/20271130

Distribution

US Nationwide distribution in the states of CA, MN, IN, MI, LA.

Quantity

1240