FDA Enforcement Class II Ongoing

Azurion systems with software release R1.x

Recall: Z-0886-2022 · Reported May 11, 2022

Enforcement

Recall Number
Z-0886-2022
Event ID
89964
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Voluntary / Mandated
FDA Mandated
Initial Notification
Letter
Report Date
May 11, 2022
Initiation Date
May 26, 2021
Classification Date
May 3, 2022
Address
Veenpluis 4-6, N/A, Best, N/A, N/A, Netherlands

Description

Azurion systems with software release R1.x

Reason

In the Azurion system, the user can add a new study to a patient by selecting the option Add Study . The Add Study dialogue box is then displayed where the Patient Type is selected to perform the study. Due to a software defect, when the study is initiated by pressing Start Procedure , the Patient Type changes inadvertently to a Patient Type different than the one selected as shown in the Table below. Patient type is one of the factors involved in the dose control process. The incorrect patient type changes the technique factors to be used by the system without notification to the user.

Code Info

software release R1.x

Distribution

US Nationwide and Worldwide Distribution

Quantity

722 systems