Azurion systems with software release R1.x
Recall: Z-0886-2022
·
Reported May 11, 2022
Enforcement
- Recall Number
- Z-0886-2022
- Event ID
- 89964
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Voluntary / Mandated
- FDA Mandated
- Initial Notification
- Letter
- Report Date
- May 11, 2022
- Initiation Date
- May 26, 2021
- Classification Date
- May 3, 2022
- Address
- Veenpluis 4-6, N/A, Best, N/A, N/A, Netherlands
Description
Azurion systems with software release R1.x
Reason
In the Azurion system, the user can add a new study to a patient by selecting the option Add Study . The Add Study dialogue box is then displayed where the Patient Type is selected to perform the study. Due to a software defect, when the study is initiated by pressing Start Procedure , the Patient Type changes inadvertently to a Patient Type different than the one selected as shown in the Table below. Patient type is one of the factors involved in the dose control process. The incorrect patient type changes the technique factors to be used by the system without notification to the user.
Code Info
software release R1.x
Distribution
US Nationwide and Worldwide Distribution
Quantity
722 systems