FDA Enforcement
Class II
Terminated
HipLOC CHS Plate/Lag Screw Introducer
Recall: Z-0885-2016
·
Reported March 2, 2016
Enforcement
- Recall Number
- Z-0885-2016
- Event ID
- 73258
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Biomet, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 2, 2016
- Initiation Date
- February 16, 2016
- Classification Date
- February 25, 2016
- Termination Date
- February 14, 2017
- Address
- 56 E Bell Dr, N/A, Warsaw, IN, 46582-6989, United States
Description
HipLOC CHS Plate/Lag Screw Introducer
Reason
A complaint was received where the Hiploc Plate/Lag Screw Introducer was undersized and would not accept the introducor coupling screw. It appears now that not all parts manufactured to Rev A drawing were reworked.
Code Info
Catalog Number: 35-000263; and Lot Number Identification: 1807188, 1807189, 1943563, and 2577847.
Distribution
US Distribution to the states of : MO, UT, AZ, IL, PA, KY, MI, NC, NJ, MN and GA., and Internationally to The Netherlands.
Quantity
23 units