Elekta, Apex add-on Micro MLC, collimator for radiation therapy.
Recall: Z-0885-2013
·
Reported March 6, 2013
Enforcement
- Recall Number
- Z-0885-2013
- Event ID
- 64213
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Elekta, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Other
- Report Date
- March 6, 2013
- Initiation Date
- January 18, 2013
- Classification Date
- February 26, 2013
- Termination Date
- May 8, 2014
- Address
- 400 Perimeter Center Ter Ne, Ste 50, Atlanta, GA, 30346-1227, United States
Description
Elekta, Apex add-on Micro MLC, collimator for radiation therapy.
Reason
If the Apex collimator is incorrectly installed on the radiation head, a touchguard inhibit activates the view restrictions dialog box only shows the top 12 restrictions.
Code Info
152652
Distribution
Nationwide Distribution including TX
Quantity
1