IMMULITE/IMMULITE 1000 Systems Prostate-Specific Antigen (PSA) Assay, REF/Catalog Number LKPTS1, Siemens Material Number (SMN) 10706279 (US); Catalog Number LKPS1, SMN 10380960 and Catalog Number LKPS5, SMN 10380949 (OUS); IVD --- Origin: UK Siemens Healthcare Diagnostics Products Ltd. Llanberis, Gwynedd, LL554EL UK For in vitro diagnostic use with the IMMULITE /IMMULITE 1000/IMMULITE 2000/2000 XPi Analyzers for the quantitative measurement of prostate-specific antigen (PSA) in human serum, as an aid in the detection of prostate cancer when used in conjunction with digital rectal examination (DRE) in men aged 50 years or older. This assay is further indicated as an adjunctive test to aid in the management of prostate cancer patients.

Recall: Z-0884-2017 · Reported December 28, 2016

Enforcement

Recall Number
Z-0884-2017
Event ID
75798
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Healthcare Diagnostics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 28, 2016
Initiation Date
November 18, 2016
Classification Date
December 22, 2016
Termination Date
June 7, 2018
Address
511 Benedict Ave, N/A, Tarrytown, NY, 10591-5005, United States

Description

IMMULITE/IMMULITE 1000 Systems Prostate-Specific Antigen (PSA) Assay, REF/Catalog Number LKPTS1, Siemens Material Number (SMN) 10706279 (US); Catalog Number LKPS1, SMN 10380960 and Catalog Number LKPS5, SMN 10380949 (OUS); IVD --- Origin: UK Siemens Healthcare Diagnostics Products Ltd. Llanberis, Gwynedd, LL554EL UK For in vitro diagnostic use with the IMMULITE /IMMULITE 1000/IMMULITE 2000/2000 XPi Analyzers for the quantitative measurement of prostate-specific antigen (PSA) in human serum, as an aid in the detection of prostate cancer when used in conjunction with digital rectal examination (DRE) in men aged 50 years or older. This assay is further indicated as an adjunctive test to aid in the management of prostate cancer patients.

Reason

Siemens Healthcare Diagnostics has confirmed that the IMMULITE/ IMMULITE 1000 Systems PSA assays are not meeting the current High-dose Hook Effect expectation in the Instructions for Use.

Code Info

All lots LKPS1/LKPS5 - Lots 443, 445, 446 LKPTS1 - Lot D116

Distribution

Worldwide Distribution-US (nationwide)and the following countries: Afghanistan, Albania, Algeria, Argentina, Aruba, Austria, Bangladesh, Belarus, Belgium, Bolivia, Bosnia Herzeg, Brazil, British Guyana, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Dominican Rep., Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Great Britain, Grenada, Guatemala, Honduras, Hungary, Iceland, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kazakhstan, Kosovo, Latvia, Lebanon, Lithuania, Macedonia, Malaysia, Mexico, Netherlands, Nicaragua, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Serbia, Slovakia, Spain, South Africa, South Korea, Suriname, Switzerland, Syria, Taiwan, Tadjikistan, Thailand, Trinidad, Tobago, Tunisia, Turkey, Turkmenistan, United Arab Emirates, Uruguay, Uzbekistan, Venezuela, and Vietnam.

Quantity

4705 kits (US: 390 kits; OUS: 4315 kits)