FDA Enforcement Class II Terminated

Philips Medical Systems, GEMINI LXL - Model #882412; GEMINI TF 16 - Model #882470 and Model #882473; GEMINI TF 64 - Model #882471, Positron Emission Tomography (PET) and X-Ray Computed Tomography (CT) scanning systems, diagnostic imaging systems for fixed or mobile installations that combine Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.

Recall: Z-0882-2013 · Reported March 6, 2013

Enforcement

Recall Number
Z-0882-2013
Event ID
63822
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Medical Systems (Cleveland) Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 6, 2013
Initiation Date
October 24, 2012
Classification Date
February 26, 2013
Termination Date
July 23, 2015
Address
595 Miner Rd, N/A, Cleveland, OH, 44143-2131, United States

Description

Philips Medical Systems, GEMINI LXL - Model #882412; GEMINI TF 16 - Model #882470 and Model #882473; GEMINI TF 64 - Model #882471, Positron Emission Tomography (PET) and X-Ray Computed Tomography (CT) scanning systems, diagnostic imaging systems for fixed or mobile installations that combine Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.

Reason

Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm whereby these generated datasets are 'flipped' and incorrectly labeled left-to-right when the are saved to disk. If the flipped Intensity Project

Code Info

The affected GEMINI systems are identified with the following Serial Numbers: GEMINI LXL Model # 882412, Serial Numbers: 14702 and 14707; GEMINI TF 16 Model #882470, Serial Numbers: 7129, 7121, 7123, 7128, 7096, 7133, 7185, 7187, 7226, and 7220; GEMINI TF 16, Model #882473, Serial Numbers: 7150, and and 7237;GEMINI TF 64, Model #882471, Serial Numbers: 7103, 7126, 7132, 7158, 7160, 7164, 7175, 7178, 7179, 7218, 7229, 7221, 7245, 7550, 7255, and 7256.

Distribution

Worldwide Distribution - USA including AK, AK, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, OA MA, MD MI, MN, MO, MS, NC, D, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV; Internationally to Canada and other countries..

Quantity

30 units