FDA Enforcement Class II Ongoing

Philips, REF M5071A, Adult. Smart PADS CARTRIDGE, 8+ years FOR USE WITH HEARSTART HS1 DEFIBRILLATORS

Recall: Z-0881-2022 · Reported April 13, 2022

Enforcement

Recall Number
Z-0881-2022
Event ID
89672
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Philips North America LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 13, 2022
Initiation Date
February 4, 2022
Classification Date
April 6, 2022
Address
22100 Bothell Everett Hwy, Bothell, WA, 98021-8431, United States

Description

Philips, REF M5071A, Adult. Smart PADS CARTRIDGE, 8+ years FOR USE WITH HEARSTART HS1 DEFIBRILLATORS

Reason

There is a potential for AED pads to experience gel separation from the foam/tin backing when peeled from the plastic liner. The gel may fold onto itself resulting in reduced surface area or gel on the pad, or it may separate almost completely leaving only a small amount of gel on the pad.

Code Info

ALL LOTS

Distribution

Worldwide distribution. US Nationwide including Puerto Rico and Guam, Germany, Australia, Sweden, Japan, Canada, Taiwan, France, Netherlands, Spain, Singapore, Austria, South Africa, Italy, United Kingdom, India, Malaysia, Switzerland, Chile, Ireland, Norway, Hong Kong, Argentina, Mexico, Bolivia, Croatia, French Guiana, Denmark, United Arab Emirates, Hungary, Belgium, Indonesia, Korea, Republic of, Bhutan, New Zealand, Bulgaria, Mauritius, Trinidad and Tobago, Israel, Guatemala, Mali, Algeria, Philippines, Malta, Curacao, Panama, Paraguay, Costa Rica, Colombia, Liechtenstein, Saudi Arabia, Monaco, Estonia, Senegal, Reunion, Luxembourg, Qatar, El Salvador, Namibia, Myanmar, Nigeria, Poland, Aruba, Greece, Martinique, Guadeloupe, Dominican Republic, China, Kosovo, Portugal, Bangladesh, Tanzania, Peru, Lithuania, New Caledonia, State of Palestine, Finland, Cambodia, Macao, Cyprus, Zambia, Viet Nam, Uruguay, Ecuador, Netherlands Antilles, Czech Republic, French Polynesia, Kenya, Cameroon, Guinea, Mayotte, and Brunei Darussalam.

Quantity

1,696,047