FDA Enforcement Class II Terminated

Abbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, potassium, ionized calcium, glucose, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary whole blood List Number: 03P88-25

Recall: Z-0880-2021 · Reported January 27, 2021

Enforcement

Recall Number
Z-0880-2021
Event ID
87073
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Abbott Point Of Care Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
January 27, 2021
Initiation Date
December 22, 2020
Classification Date
January 19, 2021
Termination Date
September 1, 2023
Address
400 College Rd E, N/A, Princeton, NJ, 08540-6607, United States

Description

Abbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, potassium, ionized calcium, glucose, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary whole blood List Number: 03P88-25

Reason

Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.

Code Info

Stored at room temperature (18 to 30¿C) for periods of time in excess of: 3 days for lot numbers *20100 to *20339 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y Primary GTIN (01) 00054749000163)

Distribution

Nationwide

Quantity

2,820,000 cartridges US and 2,273,453 catridges OUS