FDA Enforcement Class II Ongoing

LEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 71434413. (Orthopaedic manual surgical instrument).

Recall: Z-0878-2020 · Reported February 5, 2020

Enforcement

Recall Number
Z-0878-2020
Event ID
84559
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Smith & Nephew, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 5, 2020
Initiation Date
December 16, 2019
Classification Date
January 28, 2020
Address
1450 E Brooks Rd, N/A, Memphis, TN, 38116-1804, United States

Description

LEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 71434413. (Orthopaedic manual surgical instrument).

Reason

Multiple lots of LEGION AP Femoral Cutting Blocks were manufactured with an oversize green plunger hole and therefore do not meet their design specifications.

Code Info

LEGION AP Cutting Block Size 4, Product Number 71434409, Lot Numbers: 17LM05399, 17LM07852, 17LM07852A; LEGION AP Cutting Block Size 6, Product Number 71434411, Lot Numbers: 17LM05397, 17LM05398, 17LM07849; LEGION AP Cutting Block Size 7, Product Number 71434412, Lot Numbers: 17LM01746, 17LM05390, 17LM05394, 17LM07851; LEGION AP Cutting Block Size 8, Product Number 71434413, Lot Number: 17LM08442.

Distribution

USA: IA, NC, PA, TN. Int'l: CH, BE, GB, AE, CN, DE, NL, BR, FR.

Quantity

59 units