FDA Enforcement Class II Terminated

Solitaire-C Cervical Spacer System a) 14-531593 Drill/Awl Sleeve b) 14-531594 Spring-Loaded Drill/Awl Sleeve Intended to stabilize cervical spinal segment to promote fusion in order to restrict motion and decrease pain using bone graft with or without supplemental fixation.

Recall: Z-0874-2015 · Reported December 31, 2014

Enforcement

Recall Number
Z-0874-2015
Event ID
69981
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ebi, Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
December 31, 2014
Initiation Date
November 11, 2014
Classification Date
December 24, 2014
Termination Date
April 15, 2015
Address
399 Jefferson Rd, Parsippany, NJ, 07054-3707, United States

Description

Solitaire-C Cervical Spacer System a) 14-531593 Drill/Awl Sleeve b) 14-531594 Spring-Loaded Drill/Awl Sleeve Intended to stabilize cervical spinal segment to promote fusion in order to restrict motion and decrease pain using bone graft with or without supplemental fixation.

Reason

A design deficiency was discovered whereby the Drill/Awl Sleeve and Spring-Loaded Drill/Awl Sleeve instruments (the Sleeves) allow the mating Drill and Awl instruments to penetrate 1.0 mm beyond their intended depths during screw hole preparation.

Code Info

Part number 14-531593 Part number 14-531594 Catalog number Lot Number 14-531593 N23597 14-531594 N23599

Distribution

US Nationwide Distribution.

Quantity

193 units