FDA Enforcement Class II Ongoing

PageWriter TC30 w/ trolley Government Bundle, Product 860354, Software revisions up to and including A.07.05.22 to evaluate the electrocardiogram of adult and pediatric patients

Recall: Z-0859-2019 · Reported February 27, 2019

Enforcement

Recall Number
Z-0859-2019
Event ID
81957
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Philips North America, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 27, 2019
Initiation Date
January 9, 2019
Classification Date
February 21, 2019
Address
3000 Minuteman Rd, Andover, MA, 01810-1032, United States

Description

PageWriter TC30 w/ trolley Government Bundle, Product 860354, Software revisions up to and including A.07.05.22 to evaluate the electrocardiogram of adult and pediatric patients

Reason

Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to melt and/or the device to ignite, which can cause injury to patient or nearby users.

Code Info

US41510645 US41510646 US71630792 USN1206382 USN1206383 USN1409879 USO1631023

Distribution

Domestic distribution nationwide. Foreign distribution worldwide.

Quantity

7