FDA Enforcement Class II Terminated

CP5 centrifugal pump system Product Usage: The Stockert S5/ Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less.

Recall: Z-0857-2014 · Reported February 5, 2014

Enforcement

Recall Number
Z-0857-2014
Event ID
67199
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Sorin Group Deutschland GmbH
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 5, 2014
Initiation Date
December 4, 2013
Classification Date
January 28, 2014
Termination Date
December 28, 2015
Address
Lindberghstrasse 25, Munchen, N/A, Germany

Description

CP5 centrifugal pump system Product Usage: The Stockert S5/ Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less.

Reason

Sorin is initiating a field correction on S5/C5 Heart-lung machine and CP5 centrifugal pump system due to reports of pump speed control knob failures resulting in no response.

Code Info

Item Number: 60-02-60

Distribution

Worldwide Distribution - US (Nationwide) and PR; and Internationally to:BE, ES, FI, GB, NL, SE, AR, AT, AU, CA, CH, CN, CR, EG, HK, ID, JO, JP, KR, KW, LB, LY, MM, PL, RO, RS, RU, SG. TH, TN, TR, TW, UA, VN, and ZA.

Quantity

75