Orthofix Firebird Spinal Fixation System Instrumentation Modular Screw Driver, Product Usage: The Modular Screw Driver is used for inserting a modular bone screw into the pedicle during a spinal fixation procedure. It is a reusable instrument.
Enforcement
- Recall Number
- Z-0855-2014
- Event ID
- 66969
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Orthofix, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 5, 2014
- Initiation Date
- November 26, 2013
- Classification Date
- January 28, 2014
- Termination Date
- December 9, 2014
- Address
- 3451 Plano Pkwy, Lewisville, TX, 75056-9453, United States
Description
Orthofix Firebird Spinal Fixation System Instrumentation Modular Screw Driver, Product Usage: The Modular Screw Driver is used for inserting a modular bone screw into the pedicle during a spinal fixation procedure. It is a reusable instrument.
Orthofix received 6 complaints which resulted in a reportable events due to extended surgical times greater than 30 minutes for the Modular Screw Driver (PN 52-1332). The complaints alleged that the screw driver's collet would malfunction resulting in the surgeon being unable to use the Modular Screw Driver to effectively place Modular Screws, which may result in a delay of surgery.
Production Identification Numbers: Part Number 52-1332, All Lot Numbers.
Worldwide Distribution - US Nationwide including (Puerto Rico) and the countries of: Spain, Germany, Australia, Peru, Mexico, Colombia, South Africa, and Israel.