FDA Enforcement
Class II
Ongoing
Hanger-Bar 2P 45cm
Recall: Z-0852-2022
·
Reported April 6, 2022
Enforcement
- Recall Number
- Z-0852-2022
- Event ID
- 89509
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Human Care USA, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 6, 2022
- Initiation Date
- November 1, 2021
- Classification Date
- March 31, 2022
- Address
- 4210 S Industrial Dr Ste 160, N/A, Austin, TX, 78744-1077, United States
Description
Hanger-Bar 2P 45cm
Reason
The Hanger-Bar is being recalled due to the potential for the set screw loosening as a result of a manufacturing issue (failure to apply Loctite adhesive). As a result of the set screw loosening, there is the potential for a fall hazard.
Code Info
Part Number 72755 UPC Code: 881608800403 EAN Code: 7331599999714 Lot Number: 33164504
Distribution
U.S. Nationwide distribution in the states of CA, IL, LA, MA, NV, OH, OK, PA, TX, UT, VA, and WV.
Quantity
15 devices