AccuLIF PL Tubing Set and AccuLIF TL Tubing Set AccuLIF is an interbody device system that uses hydraulic pressure to expand the implant in-situ. The purpose of the AccuLIF PL Tubing Set (cat#900301) and AccuLIF TL Tubing Set (cat#905103) is to deliver pressurized saline from a syringe to the AccuLIF implant for in-situ expansion. The Tubing Set attaches to the AccuLIF Insertion handle at two district locations: the "junction block" and the "manifold." The PL and TL Tubing Sets are Class 2 instruments not intended to sustain or support life.
Enforcement
- Recall Number
- Z-0852-2016
- Event ID
- 73179
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Stryker Spine
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 2, 2016
- Initiation Date
- December 16, 2015
- Classification Date
- February 23, 2016
- Termination Date
- August 1, 2016
- Address
- 2 Pearl Ct, N/A, Allendale, NJ, 07401-1611, United States
Description
AccuLIF PL Tubing Set and AccuLIF TL Tubing Set AccuLIF is an interbody device system that uses hydraulic pressure to expand the implant in-situ. The purpose of the AccuLIF PL Tubing Set (cat#900301) and AccuLIF TL Tubing Set (cat#905103) is to deliver pressurized saline from a syringe to the AccuLIF implant for in-situ expansion. The Tubing Set attaches to the AccuLIF Insertion handle at two district locations: the "junction block" and the "manifold." The PL and TL Tubing Sets are Class 2 instruments not intended to sustain or support life.
There were two potential interference conditions identified with the way the tubing set attaches to the inserter.
Catalog number 900301, Lot #01131543, 01261508, 10151404, 10271408, 12161403, 12161404, 09021404, 01261509, 10271407, 01261507, 10071415, 05011403, 06241407, 08071402, 10091305, 03041403, 01271412, 07281401, and 11061303. Catalog number 905103, 01131542, 11241408, 12161405, 12161406, 0211501, 01211502, 11241406, 02031403, 06241408, 07161407, 09101402, 09281401, 10251303, 05301402, 11241407, and 12101305.
US Nationwide Distribution
413 units